DishFlow is an image-proposal and review tool. This page describes the deployed approach, not a claim of general biological accuracy.

What automatic counting does

The current engine uses classical computer vision, not a deployed trained neural network. It estimates the dish interior, reduces smooth illumination variation, finds bright and dark local structures, and filters connected components by size and shape. Detected barcodes and some long marker-like strokes are masked.

These are candidate image objects. Touching colonies may merge; texture, reflections or edges may look like colonies. Dense growth, unusual media, blur and glare can substantially change the result.

What a score does not mean

Image-quality and colour-consistency scores are descriptive measurements. They are not calibrated probabilities of a correct count. A clean-looking overlay or a lack of warnings does not prove accuracy. No score auto-approves a result.

What you review

On a computer, click empty space to add a mark, drag a mark to move it, and right-click a mark to remove it. On touch screens, tap to add or remove and choose Move mark for deliberate dragging. Undo and redo are available; saving requires your confirmation.

Colour groups describe visible appearance only. They do not identify an organism, prove viability, or establish a phenotype. CFU/mL uses a saved human count and explicitly entered dilution and plated volume.

Measurement boundaries

Pixels are not automatically millimetres.

Use Set image scale to mark two ends of a known reference in the same plane as the measured objects. The saved scale records its reference, image hash and revision. It assumes uniform scale; it does not correct perspective or lens distortion.

Calibrated object sizes describe automatic segmented components, not human point marks or independently validated colony boundaries. A manually added point has no measured area.

Review copies, not archival originals.

Supported inputs are still JPEG, PNG and WebP files up to 8 MB and 16 megapixels. Stored copies are oriented, stripped of metadata and resized to a long edge of at most 1,600 pixels. Keep your originals in laboratory storage.

Bring metadata once.

Optional UTF-8 CSV metadata matches every selected image exactly once by filename. It is checked before the first photo or credit is committed. Ambiguous, missing or duplicate matches are refused.

Download the metadata template

Inspect the evidence, not a marketing percentage.

The published real-image evaluation shows substantial count error. Engineering tests establish specific software behavior; they do not establish biological accuracy or time saved.